SPAG-2® User's
Guide
Treating Mild RSV at Home
Treating Severe RSV in the Hospital
Virazole as Treatment / Prescribing Information
Patient Support Materials
A. INTRODUCTION
The Small Particle Aerosol Generator Model-2 (SPAG-2) is indicated for administration
of Virazole® (Ribavirin) aerosol only. Virazole® aerosol is
indicated in the treatment of carefully selected hospitalized infants and young children
with severe lower respiratory tract infections due to respiratory syncytial virus (RSV).
B. GENERAL OPERATION OF THE SPAG-2 AEROSOL GENERATOR
The pneumatic flow system of the SPAG-2 is depicted in Figure B-1. The SPAG-2 is driven
by a pressurized gas (air or blended oxygen) source, which passes through an external
pressure compensated flowmeter before entering the SPAG-2. The incoming gas is regulated
to 26 psi (179 kPa), and then is directed, via a manifold and pressure compensated
flowmeters, to both the nebulizer and the drying chamber.
The nebulizer generates a fine aerosol of hydrated Virazole, and the drying chamber
further dehumidifies the aerosol. The mass median diameter of the aerosol particles is
approximately 1.3 microns.
C. EQUIPMENT
1 -SPAG-2.
2 - 500 ml. reservoirs (polysulfone plastic).
2 - Reservoir caps (polysulfone plastic), with gasket and 0-rings.
2 - Nebulizers.
2 - Drying chambers (polysulfone plastic).
2 - Cleanout wires for nebulizer (0.011" dia., stainless steel).
12 - Lengths nebulizer tubing (silicone).
2 - Extra nebulizer cleanout screws (stainless steel).
2 - Extra nebulizer washers (Teflon).
2 - Operator's manuals.
The spare parts included with the SPAG-2 will allow operation with little or no
downtime.
2. Operation of the SPAG-2 requires that the following equipment and items be provided
by the hospital or clinic:
2.1 Compressed gas source, regulated to 40-60 psi (pounds per square inch gauge)
(275-413 kPa). Use air which meets specifications for medical breathing use, unless the
physician prescribes oxygen which must be blended.
2.2 If greater than 21 % oxygen is to be delivered to the patient, an oxygen blender
should be used. Fraction of inspired oxygen (FiO2) should be
monitored at the point of patient delivery.
CAUTION: Oxygen therapy requires careful monitoring.
2.3 Flowmeter, pressure compensated, capable of measuring 15 liters per minute at 50
psig (344 kPa).
2.4 Hose to supply gas to SPAG-2 (equipped with a standard 9/16-18 hemispherical
connector).
2.5 Aerosol conduction tubing, 22 mm diameter.
Further equipment required will depend on the particular mode of aerosol
administration. Virazole aerosol may be delivered to a face mask, hood, tent, or
mechanical ventilator.

3. Additional equipment that must be provided by the hospital or clinic when operating
the SPAG-2 with a ventilator is listed below:
3.1 Bacteria filters
3.2 T adapter
3.3 Coupling connectors
3.4 One-way valve
3.5 Pressure monitor
3.6 Oxygen blender
3.7 Adjustable pressure relief device (Water Column Relief Valve - for volume and pressure
cycled ventilators)
IMPORTANT: Internal wire heated ventilator circuits must be used to minimize
drug precipitation.
Circuit components described in Sections F, G and H of the manual with the notation or
equivalent are generally interchangeable with items available from other manufacturers. It
is the responsibility of the user to ensure compatibility of the components and to monitor
proper operation of all components in the ventilator circuit. Equipment identification is
provided only to assist users. It is not meant as an endorsement of any company or
product.
The ICN Pharmaceuticals, Inc. Group makes no warranty for the performance of other
manufacturers equipment.
D. ASSEMBLY OF THE SPAG-2 AEROSOL GENERATOR
The following section describes the assembly of the SPAG-2 and connection to a hood,
mask, or tent. See Part 2 of this manual for instructions for use in conjunction with
mechanical ventilators.
1. Check all 3 nebulizer orifices with the cleanout wire to ensure that they are not
occluded. Insert the nebulizer stem into the swage fitting in the center of the reservoir
cap. Tighten the swage firmly with the hand tight nut.
2. Inspect the reservoir cap gasket for cleanliness. Place the cap onto the reservoir
containing the Virazole solution (see Section K for instructions describing preparation of
Virazole). Twist the cap clockwise to seal the cap onto the reservoir. Make sure that the
nebulizer pick-up tubes are not flush against the bottom of the reservoir.
3. Place the cap/reservoir assembly into the housing, and snap it into place in the
bracket.
4. Connect the drying air flow (black hose) quick-coupling to the larger fitting on the
cap spout. Connect the nebulizer air flow (blue hose) quick-coupling to the smaller
fitting on the body of the cap. Press each quick-coupling in with a twisting motion until
it snaps in place.
5. Insert the drying chamber through the side hole in the SPAG housing, and push onto
the cap spout 0-ring. The flow direction arrow on the chamber must point away from the
cap.
6. Connect the aerosol conduction tubing (supplied by the user) to the outlet port of
the drying chamber.
NOTE: The length of the aerosol conduction tubing between the SPAG-2 and
the mask, hood or tent should be kept as short as conveniently possible to minimize
collection of aerosol during operation. The tubing should be changed at the same frequency
as the wetted parts (see Section l).
7. For use with a hood, the aerosol delivery tubing from the SPAG-2 should be placed
into the inlet port of the hood.
8. For use with a face mask, a standard T piece should be connected to the mask. The
aerosol delivery tubing from the SPAG-2 should be connected to one of the open ends of the
T. Additional aerosol tubing is attached to the branch of the T for a reservoir. This
reservoir tubing is for optimal Virazole delivery. The length of the reservoir tubing will
depend on the patient's tidal volume. The practitioner should follow the hospital's
procedure for determining the patient's tidal volume. (See Appendix 1.)
9. For use with a tent, the tent must be set up with the customary air flow and cooling
systems. The aerosol delivery tubing from the SPAG-2 should be connected to the inlet port
of the tent.
E. SPAG-2 OPERATING PROCEDURE
The following section describes the general operating procedures for the SPAG-2. See
Part 2 of this manual for instructions for use in conjunction with mechanical ventilators.
NOTE: If greater than 21% oxygen is to be delivered to the patient, an
oxygen blender should be used. Fraction of inspired oxygen (FiO2)
should be monitored at the point of patient delivery.
WARNING
Whenever two or more oxygen blending systems are used, all blending systems must be
adjusted when any changes are made in FiO2. |
CAUTION: The gas source must be connected to the SPAG-2 and turned on
whenever the SPAG-2 is in use. The source must be regulated in the range 40-60 psi
(275-413 kPa). Use air which meets specifications for medical breathing use unless the
physician prescribes oxygen which must be blended. Oxygen therapy requires careful
monitoring.
1. Connect the SPAG-2 to the external flowmeter (provided by user). Open the
external flowmeter control valve completely. It is important that the external flowmeter
does not restrict gas flow to the SPAG-2.
2. Open the nebulizer flowmeter valve completely (approximately six turns counter
clockwise from the closed position).
3. Disengage the lock sleeve on the SPAG-2 pressure regulator by pulling the adjustment
knob outward. For SPAG-2 devices with a white lock ring on the pressure regulator,
disengage the lock ring by pulling the ring outward. Adjust the pressure regulator knob
(clockwise to increase) until 26 psi (179 kPa) is indicated on the SPAG-2 pressure gauge.
4. Open the drying air flowmeter valve from the closed position until the external
flowmeter indicates the desired total flow through the unit. It is suggested that the
external flowmeter should read approximately 15 liters per minute (LPM) when using a hood
or a tent, and approximately 12 LPM when using a face mask.
5. Readjust the SPAG-2 pressure regulator to 26 psi (179 kPa).
6. Verify that the external flowmeter reading is correct and, if necessary, adjust
again by turning the drying air flowmeter valve.
7. The SPAG-2 is correctly adjusted when the operating parameters stabilize into the
following ranges:
7.1 Regulator pressure = 26 ± 2psig
(165-193 kPa).
7.2 Nebulizer flowmeter range from 6-10 LPM.
7.3 Drying air flowmeter = range from 2-9 LPM.
Read all rates of flow at the center of the indicator ball.
.
8. TROUBLESHOOTING
8.1 INCORRECT AIR FLOW READING: Ensure that the SPAG-2 flows have stabilized
before deciding that they are out of tolerance.
a. Nebulizer air flow: If the indicated nebulizer air flow is greater than the
specification in 7.2 above, check for leaks in the connections delivering air to the
nebulizer. Ensure that the clean out screw in the nebulizer is not leaking. If the flow is
less than the specification in 7.2, clean the nebulizer and check for occlusions in the
orifices. If these measures do not bring the flow into the correct range, change the
nebulizer.
b. Drying chamber air flow: Ensure that the flow to the nebulizer is within the
specified range before troubleshooting the drying chamber air flow. If the flow to the
drying chamber cannot be adjusted within the specified range, check for leaks or
occlusions in the plumbing connections.
8.2 NEBULlZER FUNCTION
Check that all three (3) orifices of the nebulizer are
spraying correctly by looking for three approximately one-half inch diameter spray spots
on the wall of the flask. If any problem is detected, clean or change the nebulizer.
The SPAG-2 should nebulize the drug solution from the reservoir at a rate of 12.5-15.0
ml/hour. If the nebulization rate is slow, ensure that the external flowmeter (provided by
the user) is completely opened, and that the nebulizer orifices are not clogged. Change
the silicone tubing or the nebulizer tubing assembly after ensuring that other operating
parameters are correct. If the nebulization rate appears to be fast, check for excessive
leaks at the cap gasket.
9. Monitor the operation of the SPAG-2 for approximately ten minutes and adjust any
settings accordingly.
CAUTION: During operation, water/drug collection may form in the reservoir
cap spout, inside the drying chamber, at the outlet port of the drying chamber and in the
22 mm aerosol conduction tubing. Monitor these components hourly. Collection may increase
after 10 hours of operation. If excessive collection is seen, these components should be
cleaned or replaced.
10. Before inactivating the aerosol generator, assure that the patient has been
removed from the aerosol and has an alternative source of oxygen or air. To inactivate
the aerosol generator, turn off the gas source.
INSTRUCTIONS FOR USE OF THE SPAG-2 IN
CONJUNCTION WITH MECHANICAL VENTILATORS
WARNING:
Virazole® Aerosol may be used for infants requiring assisted
ventilation, but requires careful monitoring by experienced hospital personnel. Drug
precipitation in the system can be a serious problem. Mechanical ventilators used in
conjunction with a SPAG-2 should utilize an internal wire heated ventilator circuit to
minimize accumulation of drug precipitate. In addition, bacteria filters in the expiratory
limb of the circuit and a high PEEP alarm should be utilized. A water column relief valve
must be employed in circuits with volume-cycled ventilators and may be used as well as in
circuits with pressure cycled ventilators. |
F. OPERATION OF THE SPAG-2 WITH A VENTILATOR-GENERAL INSTRUCTIONS:
The following instructions apply to use of the SPAG-2 with either pressure cycled or
volume cycled ventilators and must be read before continuing.
Also see additional specific instructions for volume cycled ventilators (Section G), or
pressure cycled ventilators (Section H).
1. Use an oxygen blender as the gas source if greater than 21 % oxygen is to be
delivered to the patient. Fraction of inspired oxygen (FiO2) should
be monitored at the point of patient delivery.
WARNING:
Whenever two or more oxygen blending systems are used, all blending systems must be
adjusted when any changes are made in FiO2. |
2. Turn off the drying air flowmeter on the SPAG-2 unit.
3. Adjust the nebulizer flow to 6-10 LPM.
4. Connect the SPAG-2 delivery tube into the inspiratory line of the patient breathing
circuit immediately downstream from the heated humidifier. Place a one-way valve in the
SPAG-2 tubing at the junction with the ventilator circuit. (See Figure F-1). Check the
operation of the one-way valve before installing it into the circuit.

Figure Fl. Junction of SPAG-2 with ventilator circuit, and location of
bacteria filters
CAUTION: Water/drug precipitate may form in the one way valve, the
ventilation tubing and/or the endotracheal tube; frequent (hourly) monitoring of these
components for precipitate is required. If the valve and/or tubing collect precipitate,
they must be cleaned or replaced.
IMPORTANT: If airway pressures increase, the endotracheal tube should be
checked, and the patient suctioned, to minimize precipitate build-up. Airway pressure
increases can also be caused by clogged bacteria filters (see Item 5 of this section).
5. Bacteria filter(s) (Pall BB-50T, Marquest' #MQ-303-01, or equivalent) are mandatory
in the expiratory line just before the ventilator exhalation valve entrance port (see
Figure F-1). Filters should be changed every 2-4 hours, or more frequently as needed.
CAUTION: Filters should be changed if an increase in peak inspiratory
pressure (PIP) or positive end expiratory pressure (PEEP) of 1-2 centimeters of water is
observed. Because these pressure increases may occur, airway and ventilator circuits
should include CONTINUOUS MONITORING as described in item 8 of this section.
NOTE: Two filters may be placed in series. When changing filters, discard
the upstream filter. Place a fresh filter downstream of the remaining filter.
6. A bacteria filter may be placed in the inspiratory line upstream of the humidifier
to prevent Virazole from entering the ventilator, should a reversal of flow occur during
operation.
7. Connection of the SPAG-2 to the ventilator circuit may cause the end expiratory
pressure to increase (1-2 cm H2O). To maintain the desired positive end
expiratory pressure (PEEP) level, adjust the ventilator PEEP control.
8. Use of a pressure monitor with both high and low PEEP alarms is essential to assist
the user in detecting changes in the PEEP level (e.g. Novametrix pneumogard #1250, Bunnell
#00120, or equivalent).
9. Water condensation (rainout) associated with the use of conventional tubing may
require frequent (hourly) removal of the condensate. To minimize rainout, internal
wire-heated tubing must be used.
10. Maintain the temperature in the circuit at 33-37°C. It may be necessary to set the
heated humidifier at a higher temperature. BEFORE DlSCONNECTlNG THE SPAG-2 FROM THE
VENTILATOR CIRCUIT, THE TEMPERATURE AT THE HEATED HUMIDIFIER MUST BE ADJUSTED
APPROPRIATELY.
CAUTION: Both the heater/humidifier and the internal wire heater sensors must
be checked hourly for evidence of water/drug precipitate which can decrease the accuracy
of the sensor. When internal wire heated circuits are used, precipitate may also form at
the loop of the wire.
10.1 When using conventional tubing, place the heated humidifier temperature sensor at
the patient wye.

10.2 When using internal wire heated tubing, place the heated humidifier temperature
sensor immediately downstream from the junction of the SPAG-2 tubing and the inspiration
line of the ventilator. Place the internal wire heater temperature sensor at the patient
wye. Set the internal wire heater to 37°C, and the heated humidifier slightly less
(approximately 33°C) to maximize prevention of rainout.

11. Both the SPAG-2 tubing and the conventional ventilator circuits should be changed
daily. It is recommended that internal wire heated circuits be changed every 8-10 hours.
G. OPERATION OF THE SPAG-2 WITH A VENTILATOR - ADDITIONAL INSTRUCTIONS FOR VOLUME
CYCLED VENTILATORS
1. Read Section F before continuing with this section.
2. Figure G-1 is a schematic drawing of the ventilator tubing circuit used with the
SPAG-2 in conjunction with a volume cycled ventilator. Pediatric clinical experience with
the Siemens Servo 900C suggests that any equivalent volume cycled ventilator may be used
with the SPAG-2.
Figure G-1.Schematic of ventilator and SPAG-2 system used with a
volume cycled machine.
(A): Silicone rubber coupling (Puritan-Bennett #3443, or equivalent).
(B): Disposable aerosol T adapter (Hudson #1077, or equivalent).
(C): Bacteria filters (Marquest #MQ-303-01, Pall #BB-50T, or equivalent).
(D): One-way valve (Inspiron #001671, or equivalent).
(E): ICN Water Column Relief Valve (ICN #6003, see replacement parts section of this
manual).
3. The calculated volume of the tubing between the one-way valve and the patient wye
must be at least one tidal volume for correct aerosol delivery. Figure G-2 below indicates
the section of tubing of concern. See Appendix 1 as an aid to estimate tidal volume and
tubing volume.
Figure G-2. Section of ventilator circuit tubing where one patient
tidal volume is required.
4. Exhaled volumes displayed on the ventilator will read falsely high due to the
additional volume (flow) supplied by the SPAG-2.
5. Use of a low-pressure monitor (e.g. Ventronics® Model 5522 or
equivalent) will assist the user in detecting potentially dangerous low pressure
situations.
CAUTION: Certain alarm settings, such as the low minute volume alarm on the
Siemens Servo 900C, will read falsely high and thus rendered unreliable when the SPAG-2 is
connected to the ventilator.
6. Attach a water column pressure relief valve to the SPAG-2 output with a T connector
(see Figure G-1). The pressure relief valve is needed to vent the continuous flow from the
SPAG-2 during the inspiratory phase of a volume cycled ventilator.
An ICN Water Column Relief Valve is available with mounting brackets for the SPAG-2
(ICN #6003, see replacement parts section of this manual). A user supplied water column
relief valve may also be used.
Adjust the relieving pressure between 2-5 cm of water greater than the observed end
expiratory pressure (EEP) prevailing in the patient/ventilator circuit. When adjusted
properly the relief valve will bubble during the inspiratory phase of the ventilator and
cease bubbling during the expiratory phase. When using the ICN Water Column Relief Valve,
thoroughly read the instructions for use provided with the valve.
CAUTION: Monitor the relief valve hourly to ensure proper operation. Adjust
the water level in the relief valve according to any changes made in the EEP during the
course of therapy. Maintain the water level in the relief valve at 2-5 cm H2O
greater than the prevailing EEP. Clean or replace the valve as necessary.
WARNING:
Failure to place a one-way valve at the junction of the SPAG-2 tubing line and the
inspiratory limb of the ventilator will result in reducing the tidal volume delivery by
diverting the ventilator's output through the pressure relief valve. |
7. TROUBLESHOOTlNG RELIEF VALVE OPERATION:
7.1 Relief valve bubbles continuously:
a. Relief valve pressure is set too low. Increase water level in reservoir to 2-5 cm H2O
greater than the prevailing EEP level.
b. One-way valve is stuck closed or clogged. Clean or replace the< one-way valve.
7.2 Relief valve never bubbles:
a. Relief valve pressure is set too high. Decrease water level in reservoir to 2-5 cm H2O
greater than the prevailing EEP level.
b. One-way valve is stuck open. Clean or replace the one-way valve.
H. OPERATION OF THE SPAG-2 WITH A VENTILATOR ADDITIONAL INSTRUCTIONS FOR PRESSURE
CYCLED VENTILATORS
1.Read Section F before continuing with this section. in treating very small infants
with a pressure cycled ventilator, the flow rate may be considerably less than 6 LPM which
makes it impossible to employ the method of Virazole administration described in this
section. A Water-Column Relief Valve can be inserted in the SPAG circuit, just as would be
done with a volume cycled machine (see Section G). This modification is advantageous
because the controls of the pressure cycled ventilator need not be altered. In addition,
ventilator adjustments need not be made during periods when SPAG flow must be discontinued
(such as in preparation for suctioning the patient).
2.Figure H-1 is a schematic drawing of the ventilator tubing circuit used with the
SPAG-2 in conjunction with a pressure cycled ventilator. Any of various ventilators, such
as the Bear® BP-200, Healthdyne®- 105, Sechrist® Model lV-100B, or equivalent, can be
used in assisted ventilation, controlled ventilation, intermittent mandatory ventilation
(IMV), or continuous positive airway pressure (CPAP) mode.
Figure H-I. Schematic of ventilator and system used with a pressure
cycled machine.
(A): Silicone rubber coupling (Puritan-Bennett® #3443, or equivalent).
(B): Disposable aerosol "T" adapter (Hudson® #1077, or equivalent).
(C): Bacteria filters (Marquest® #MQ-303-01, Pall® #BB-50T, or equivalent).
(D): One-way valve (lnspiron® #0016671, or equivalent).
(E):ION Water Column Relief Valve (lCN #6003, see replacement parts section of this
manual).
3.Adjust the ventilator flow setting so that the peak inspiratory pressure (PIP)
observed is the same as before the SPAG-2 was connected to the circuit.
GENERAL INFORMATION
I.STERILIZATION OF SPAG-2 WETTED COMPONENTS
All parts of the SPAG-2 are clean as delivered. It is the responsibility of the user to
sterilize all wetted parts before beginning each course of therapy. At a minimum,
sterilization of all wetted parts should be done daily.
Sterilize, by autoclaving, the following parts:
1. Drying chamber (polysulfone plastic).
2. Reservoir (polysulfone plastic).
3. Reservoir cap (polysulfone plastic) with gasket and 0-rings.
4. Nebulizer assembly with tubing.
5. ICN Water Column Relief Valve, with plug, if used.
Autoclave these items by standard procedures.
Hospital personnel should follow their established procedures for monitoring bacterial
contamination of respiratory equipment for the care of the SPAG-2.
J. CLEANING OF COMPONENTS
1. Any Virazole solution remaining in the reservoir flask after treatment should be
discarded.
2. Virazole is a very water soluble material. The components (reservoir, cap,
nebulizer, drying chamber, and water column relief valve) can be effectively cleaned with
warm tap water and mild detergent. After cleaning, rinse with warm tap water followed with
distilled or deionized water.
3. Extra attention should be given to cleaning the nebulizer. Do not disassemble
unless an occlusion is suspected. Probe the orifices of the nebulizer with the
0.011" cleanout wire to dislodge any material. Ensure that the wire is inserted
completely into each of the three small inner holes on the side of the nebulizer tip.
After cleaning and rinsing the nebulizer, shake it to remove water from the small
passageways.
4. If disassembly of the nebulizer becomes necessary, remove the small cleanout screw
(with washer) and the silicone tubing. Probe all orifices with the cleanout wire. Soak all
nebulizer parts in warm to hot water containing detergent, rinse thoroughly, then blow dry
with compressed air. Reassemble the nebulizer.
5. The reservoir, cap, nebulizer, drying chamber, and water column relief valve should
be sterilized by autoclaving. (See Section I.)
6. The housing can be cleaned with any disinfectant currently used by the hospital for
cleaning the exterior of similar respiratory therapy equipment (e.g., ventilators, IPPB
devices). For example, Sanamaster III (quaternary compound), 1 oz/gallon.
K. PREPARATION OF VlRAZOLE® FOR USE IN THE SPAG-2
Virazole is supplied as 6 grams of lyophilized drug per 100 ml vial for aerosol
administration only. By sterile technique, dissolve the drug with sterile USP water for
injection or inhalation in the 100 ml vial. Transfer to the clean, sterilized 500 ml
SPAG-2 reservoir and further dilute to a final volume of 300 ml with sterile USP water for
injection or inhalation. The final concentration should be 20 mg/ml.
Important: This water should not have had any antimicrobial agent or other
substance added. The solution should be inspected visually for particulate matter and
discoloration prior to administration. Solutions that have been placed in the SPAG-2 unit
should be discarded at least every 24 hours and when the liquid level is low before adding
newly reconstituted solution.
Using the recommended drug concentration of 20 mg/ml Virazole as the starting solution
in the drug reservoir of the SPAG-2 unit, the average aerosol concentrations for a 12 hour
period would be 190 micrograms/liter (0.19 mg/l) of air.
Vials containing the lyophilized drug powder should be stored in a dry place at 15-
25°C (59-78°F). Reconstituted solutions may be stored, under sterile conditions, at room
temperature (20-30°C, 68-86°F) for 24 hours. Solutions which have been placed in the
SPAG-2 unit should be discarded at least every 24 hours.
NOTE: Levels of Virazole in the environment are expected to be very low
during proper administration of the drug. To help minimize incidental exposure, Virazole
should be administered in a well-ventilated hospital setting. Hospital staff responsible
for aerosol delivery should consider using a protective face cover (nose and mouth) to
reduce incidental exposure. Common surgical masks do not provide a significant amount of
protection. The 3M Model 9970 personal mask has been identified as one which supplies
significant protection when properly fitted. NIOSH has recommended: Training programs
should be developed; scavenging systems should be used, where possible, to reduce
environmental exposure; treatment should be administered in rooms with at least six air
changes per hour; treatment rooms should be under negative pressure; where possible, turn
off the SPAG-2 five minutes before each health care worker enters the room; avoid dermal,
oral, or ocular contact with airborne Virazole; respirators may be used to further reduce
Virazole exposure; American Academy of Pediatrics recommendation for patient selection
should be considered.
L. SELECTED REPLACEMENT PARTS FOR THE SPAG-2
Replacement parts and technical assistance for the SPAG-2 6000 Series may be obtained
by contacting your local representative or the ICN office listed below. When ordering,
please specify the SPAG-2 device serial number, as well as the part number and description
of the item desired.
Identification of any item(s) of adjunct equipment has been performed for the sole and
express purpose of assisting clinicians in the administration of Virazole with the SPAG-2.
Such identification or designation is not meant to constitute and should not be construed
as an endorsement of any company or product.
The ICN Pharmaceuticals, Inc. Group makes no warranty or representation, either express
or implied, with respect to any other manufacturer's items or equipment, their quality,
performance, or fitness for a particular purpose. In no event will the ICN
Pharmaceuticals, Inc. Group be liable for direct, indirect, special, incidental or
consequential damages resulting from any defect in construction of performance of other
manufacturer's equipment.
ICN Pharmaceuticals, Inc.